HDL-C HDL-C holesterol assay kit (Direct method)
In vitro test for the quantitative determination of HDL-Cholesterol (HDL-C) concentration in serum on photometric systems.HDL cholesterol is inversely related to the risk of developing coronary artery disease. A low HDL/LDL cholesterol ratio is directly related to the risk of developing coronary artery disease. A high HDL cholesterol is associated with the “longevity” syndrome.
Storage and stability
Up to expiration date indicated on the label, when stored unopened at 2-8℃ and protected from light.
Once opened, the reagents are stable for 28 days when refrigerated on the analyzer or refrigerator.
Contamination of the reagents must be avoided.
Do not freeze the reagents.
Once dissolved, the calibrator are stable for 7 days at 2–8℃, the control are stable for 7 days at 2–8℃ , do not freeze.
Principle
The System monitors the change in absorbance at 546nm. This change in absorbance is directly proportional to the concentration of cholesterol in the sample and is used by the System to calculate and express the HDL-cholesterol concentration.
Reagents preparation
Component |
Concentrations |
Reagents 1(R1): |
|
Good’s Buffer |
100mmol/L(PH7.5) |
Polyanion |
0.5mmo1/L |
CE |
800U/L |
COX |
500U/L |
POD |
30KU/L |
4-AAP |
0.5mmo1/L |
Reagents 2(R2): |
|
Good’s Buffer |
100mmol/L(PH7.0) |
TOOS |
1mmol/L |
Surfactant |
Right amount |
Sample requirements
Serum is suitable for samples. Whole blood, hemolysis and urine are not recommended for use as a sample. Freshly drawn serum is preferred specimen.
Calibration Preparation
Carefully open the bottle, avoiding the loss of lyophilizate, and pipette in exactly 1.0 mL of distilled/deionized water. Carefully close the bottle and dissolve the contents completely by occasional gentle swirling within 30 minutes. Avoid the formation of foam. The dissolved calibrator can be used without any other Pretreatment.
Method
1. Reagent preparation :Liquid reagent can be used when opened.
2. Measurement:
Main wavelength 546nm
Sub wavelength 700nm
Temperature 37℃
Type 2-Endpoint
Calibration
It is recommended to use the Calibrator from Hyasen and distilled/deionized water for two-point calibration.
Calibration frequency:
After reagent lot changed.
As required following quality control procedures.
Quality control
At least two levels of control material should be analyzed with each batch of samples. Each laboratory should establish
its own internal quality control scheme and procedures for corrective action if controls do not recover within the acceptable tolerances.
Sample (calibration) |
3μL |
R1 |
225μL |
Mix and keep at 37℃, 5 min, determination of absorbance A1. |
|
R2 |
75μL |
Mix and keep at 37℃, 5 min, determination of absorbance A2. |
△A=[(A2-A1)sample / calibrator]-[(A2-A1) blank]
Reference intervals [1,2]
Each laboratory should establish its own reference intervals based upon its patient population. The reference intervals measured at 37℃ listed below were taken from literature:
Man>0.9mmol/L
Female>1. 15mmol/L
Interferences/specificity
The following substances were tested for interference with this methodology. Criterion: Recovery within ±10 % of initial value.
Bilirubin <50 mg/dL, intralipid <3000 mg/dL, ascorbic acid <50 mg/dL, hemoglobin <500 mg/dL
Severe lipemia and hemolysis will affect the measurement results. The components in the kits of different batch numbers cannot be exchanged or mixed.
Reagent blank absorbance
The absorbance of reagent blank at 546nm should be <0.03.
Sensitivity/detection limit
When testing 1.00mmol/L HDL cholesterol, the absolute value of absorbance difference should not be less than 0.04.
Precision
Within-run : CV≤3%
Between-run: CV≤5%
Linearity range
In the interval of 0. 15 ~ 6.00mmol/L, the correlation coefficient r should not be less than 0.990; In the interval of [0. 15 ~ 2.50]mmol/L, the absolute deviation between the measured concentration and the estimated value should not exceed ± 0.25mmol/L; In the interval of (2.50 ~6.00]mmol/L, the relative deviation between the measured concentration and the estimated value should not exceed ±10%.
Warnings and precautions
1. For in vitro diagnostic use.
2. Take the necessary precautions for the use of laboratory reagents.
3. Preservative contained. Do not swallow. Avoid contact with skin and mucous membranes.
4. Disposal of all waste material should be in accordance with local guidelines.
References
[1] National Committee for Clinical Laboratory Standards, How to Define, Determine, and Utilize Reference Intervals in the Clinical Laboratory, Approved Guideline, CLSI publication C28-A, Villanova, PA(1995).
[2] Tietz, N. W., ed., Fundamentals of Clinical Chemistry, 3rd Edition, W.B. Saunders, Philadelphia, PA (1987).
[3] CLSI. Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition. CLSI document EP5-A2 [ISBN 1-56238-542-9. CLSI, 940 West Valley Road, Suite 1400, Wayne, PA19087 USA, 2008.963-968.