APS-5 Chemiluminescent Substrate Solution
APS-5chemiluminescent substrate solution is a new generation of immunodetection chemiluminescence Optical liquid, this product is a water-based mixed solution based on APS-5 compound liquid, through the optimization of key components such as enhancers, stabilizers, and preservatives, The performance of the product has been greatly improved, and it can be used in the action of alkaline phosphatase (ALP). A chemiluminescence reaction occurs under use, and photons are rapidly released. After adding the sample The maximum luminous value can be reached in about 2 minutes, and the plateau period lasts long. This product is in the range of ALP enzyme concentration of 10-4-10-8 U, The number of photons released is proportional to the concentration of ALP in the solution, suitable for Chemiluminescence immunodetection reagent system with ALP as labeling enzyme, with It has the advantages of fast response, high sensitivity, high stability, and high signal-to-noise ratio.
Product Information
Name |
APS-5 Chemiluminescence Substrate Solution |
Appearance |
Light yellow clear liquid |
Storage condition |
2-8℃, stricktly protected from light |
Operation temperature |
22-35℃ |
Stability |
2-8℃, stable for more than 1 year |
Package |
Brown plastic bottle |
Recommended dosage |
100-200 μL per test |
Applications
APS-5 is for alkaline phosphatase (Alkaline Phosphatase, ALP) as a standard Chemiluminescent immunodetection of markers. Suitable for tube luminescence, plate luminescence, POCT chemiluminescence detection, etc. same scene.
Storage Conditions
Transported with ice bag, stored at 2-8℃ for more than 1 years. strictly protected from light
Features
1. The background value is low, the luminous value is high, the signal-to-noise ratio is high, and the maximum luminous value is reached. The time is short, the ALP luminescence reaction enters the plateau phase within 2 min, and The luminescence value can remain stable for a long time (Fig. 1).
2. High sensitivity, can detect 10 -8 U or even lower concentration of ALP enzyme son.
3. Wide linear range, ALP concentration is 5 orders of magnitude of 10-4-10-8U The luminescence values in the range are linearly related (Figure 2)
4. The validity period is long, and it can be stored at 2-8℃ in the dark, and can be stored stably for more than one year.
Notes
1. It is normal for crystals to appear at the bottom of the refrigerated bottle of this product, and it should be balanced to room temperature. Warm (22-35℃) crystals disappear after use.
2. Strictly avoid light, under any circumstances (including before use) should avoid exposure to strong light, short time exposure will cause background signal to show.
3. Do not freeze.